The SMART PFM is a distinctive analyser that uses microscopy to measure the oil content in a sample and is approved to MEPC.107(49).

SMART PFM 107 OIL-IN-WATER MONITOR

Related Product Information

The monitor measures particulates in the sample stream on a continuous basis by passing the process fluid through a proprietary photo optical measuring cell developed by Rivertrace.

Using a combination of optical recognition algorithms and light intensity it is possible to differentiate between Oil particles, Gas/Air Bubbles and Solid particulates in the range 1-500 microns. Flow and particulate characteristics can be visualised live via the 7” display.

Oil concentration, pressure, temperature and oil alarm status are displayed on an HD touchscreen. Oil concentrations, alarms and fault log are stored within the system to comply with the reporting requirements of IMO resolution MEPC.107(49) and can be downloaded onto a pc via LAN or USB for further analysis.

The SMART PFM 107 offers a choice of auto cleaning methods to ensure the accuracy is maintained at all times. The cleaning is fully automatic and operates whenever it senses contamination of the optical windows.

  • Oil type independent
  • Solid discrimination to MEPC.107(49)
  • Uses Microscopy to measure oil content
  • Automatic cleaning of the optical cell
  • 7″ full colour touch screen.
  • Camera display to view contaminants within the sample.

  • Drill Rig Slop Tanks
  • Oily Water Separator Discharge
  • Regulatory Compliance
  • Produced Water Discharge

  • Hazardous Environments (Zone 1 & 2) – The SMART PFM can be offered in an explosion proof cabinet
  • Sample Conditioning – the SMART PFM can be packaged as a system with sample conditioning equipment to suit the application
  • Marine and Industrial versions of the SMART PFM can be offered. The industrial version allows for variable alarm points

SMART PFM safe area version

SMART PFM Ex version

IMG 7565 no background NEW

SPECIFICATION

MEASUREMENT 
Oil concentration range0 – 40 ppm
Oil concentration accuracy+/- 1 ppm
Oil alarm 1 operating point15 ppm standard – (1-15 ppm adjustable)
Oil alarm 2 operating point15ppm Standard – (1-40 ppm adjustable)
Oil alarm 1 operating delay1 – 2 seconds
Oil alarm 2 operating delay5 – 70 seconds
Alarm contact rating5 Amps at 230 VAC
Alarm relay modeDe-energised in alarm state
Output signal4-20 mA
Network communicationEthernet (RJ45)
Wi-Fi optional with additional hardware
IMO Data transferUSB 2 Memory stick
Ambient temperature0°C to 55°C
Sample temperature1 to 60°C
Sample flowrate0.5 to 2 l/min
Sample pressure0.5 – 6 bars
Clean water requirementsRequired
Auto Clean PressureCompressed air at 4 – 6 bar
Weight29.5 kg (65 lbs)
Standard Dimensions [mm]Height: 528mm (20.8”), Width: 600mm (23.6”), Depth: 280mm (11”)
Supply voltage100-240 VAC
Supply variation+/- 10% of nominal voltage
Supply frequency50 – 60 Hz
Power consumption50 VA normal, 100 VA peak
Degree of ingress protectionIP65
ApprovalsMED & MEPC.107 (49)
Display7″ LCD Display Screen

Specifications and system descriptions accurate at time of publishing. These are subject to change.