A versatile and sophisticated monitor to suit multiple applications in marine and industrial environments. The Smart 50M uses a nephelometry based detection technique and uses sophisticated algorithms to detect Oil in the sample
The design of the measuring cell allows the monitoring of high temperature and high pressure samples across multiple ranges of Oil Concentration. Auto-Clean and Auto-Zeroing functionality ensures the maintenance is minimal and that the measuring cell is kept free from fouling.
The Smart 50M can provide the Oil Concentration value to a DCS or PLC using a 4-20mA output.
As standard the following data is available:
- Oil Concentration
- Sample Pressure
- Sample Temperature
- Fault Status
- Alarm and Run states
Applications
- Hydrocarbon Leak Detection
- Lube Oil Cooler Monitoring
- Boiler Feed Protection
Heat Exchanger Leak Detection
Industrial Wastewater/Groundwater discharge
SPECIFICATION
Ranges: | Low range 0 – 10ppm
Medium range 0 – 200ppm |
Accuracy: | Low range +/- 0.5ppm
Medium range +/- 5 ppm |
Alarm 1 Operating Point: | Low range 1 – 10 ppm
Medium range 1 – 200 ppm |
Alarm 2 Operating Point:
|
Low range 1 – 10 ppm
Medium range 1 – 200 ppm |
Alarm 1 Operating Delay:
|
1 – 15 seconds |
Alarm 2 Operating Delay:
|
1 – 600 seconds |
Alarm contact rating:
|
8 AMP @230VAC |
Output signal:
|
4-20 mA |
Ambient temperature:
|
1-60 °C / 34 – 122 F |
Humidity:
|
Max 98% non condensing |
Sample temperature:
|
1 – 95 °C / 34-203 F |
Sample Flow:
|
0.3 – 3 Ltrs per min / 0.079 – 0.79 Gal per min |
Sample Pressure: | 1.0 Bar Minimum / 14.50 Psi
3.0 Bar Nominal / 43.50 Psi |
Clean Water requirements: | 0.3 – 3L per min / 0.079 – 0.79 Gal per min for ‘zeroing’ |
Weight: | 15kg / 33.07 lb |
Size:
|
500 x 360 x 150mm / 19.7” x 14.2” x 5.9” |
Supply voltage:
|
115 – 230 VAC |
Power consumption:
|
15 VA |
Degree of Protection:
|
IP56 |
Approvals:
|
Tested to EN61010 EN61326 (CE), c UL us. |
Specifications and system descriptions accurate at time of publishing. These are subject to change.